New Era NE-9000 Series Programmable Peristaltic User Manual

Page 40

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-9000

Publication #1200-04

32

Phase

Function

Rate

Volume

Direction

8

DECR

1.0

0.1 ml

Dispense

Phase

Function

9

LP:ST

Phase

Function

Rate

Volume

Direction

10

INCR

1.0

0.1 ml

Dispense

Phase

Function

11

LP:50

Phase

Function

12

JP:02

11.4.4

Example 4: Complex Dispenses with External Synchronization

0

0

750

W

it

h

d

ra

w

(

m

l/

h

r)

I

n

fu

s

e

Start
Trigger

Start
Trigger

:60

:60

:60

Start
Trigger

Start over

Syringe

Refill

A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate
during the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the
pump waits for a start trigger before starting the next dispense.

The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set
of dispenses, the syringe is refilled by the amount dispensed, 17.25 ml. Then the buzzer beeps, to alert the
operator to the start of the first set of dispenses. The process is then repeated.

Phase

Function

Rate

Volume

Direction

1

RATE

750.0 ml/min

0.5 ml

Dispense

Phase

Function

Rate

Volume

Direction

2

RATE

300.0 ml/min

1.5 ml

Dispense

Phase

Function

3

BEEP

Phase

Function

4

PS:00

Phase

Function

5

LP:02

Phase

Function

Rate

Volume

Direction

6

RATE

750.0 ml/min

0.5 ml

Dispense

Phase

Function

Rate

Volume

Direction

7

RATE

300.0 ml/min

1.5 ml

Dispense

Phase

Function

8

BEEP

Advertising