New Era Pump Systems Inc.

www.SyringePump.com

Model NE-9000

Publication #1200-04

21

10.3

Dispense, Pause, Repeat Mode

n

Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When enabled, will create a more automated dispensing system, whereby the pump will continuously
repeat the current dispense after a fixed time delay. The time delay will be displayed next:

nn

n.n

Where ‘nn’ is the pause time in seconds. Use the arrow keys to change the pause time. Valid pause times
are 1 to 99 or 0.1 to 9.9 seconds. To set/clear the decimal point, scroll the right-most up arrow key past 9,
or simultaneously press the 2 right-most up arrow keys.

10.4

Advanced Settings Menu

The “Other” menu selection provides access to the more advanced settings menu.

10.5

Power Failure Mode

n

Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When enabled, if the Pumping Program was operating when power to the pump was disrupted, the
Pumping Program will automatically start operating when power is reconnected to the pump. Pressing
any key on the keypad while powering up the pump will stop the Pumping Program from starting.

In Dispense Mode, the pump will begin pumping according to the last selected Dispense.

CAUTION: If Dispense Mode is not set, then Pumping Program will start operating from the

beginning of the Pumping Program (Phase 1), regardless of what part of the Pumping Program was
operating when the power was disrupted.

10.6

Audible Alarm Enable

n

Setting: ‘0’ = Disabled, ‘1’ = Enabled.

When alarms are enabled, the buzzer will be sounded as follows:

Condition

Buzzer Action

Pumping Program ended

Continuous beeping

Pumping Program paused for start trigger

Continuous beeping

Alarm condition

Steady alarm

Pressing any key will stop the alarm.

10.7

TTL I/O Operational Trigger Configuration

aa

Configures how the TTL I/O ‘Operational Trigger’ (pin 2) will control the Pumping

Program’s operation. (See sec. 13.1, TTL I/O Operational Controls). The 2 letter configuration Setting
parameter to the right of the colon (:) is defined as follows: