New Era NE-9000 Series Programmable Peristaltic User Manual

Page 35

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-9000

Publication #1200-04

27

11.3.3

‘DECr’: Decrement Rate Function

The decrement function subtracts the specified rate from the current pumping rate. For a full

description, see section 11.3.2, ‘InCr’: Increment Rate Function

11.3.4

‘StoP’: Stop Pumping Operation and End the Program

Stops the pumping operation and stops the Pumping Program. The Pumping Program will

begin at Phase 1 when started again. An implicit ‘Stop’ function is executed when the Program exceeds
the maximum Phase number during operation.

If alarms are enabled, the buzzer will beep continuously when the Pumping Program stops.

11.3.5

‘JP:nn’: Jump to Phase

nn

The ‘Jump’ function alters the consecutive operation of Program Phases. When executed, the

Pumping Program will continue operation with Phase ‘nn’.

11.3.6

'Pr:In': Program Selection Input

The Pumping Program can be broken into sub-programs which can be selected by the user.

Auto-Run Production Mode: When Phase 1 is set to Program Selection Input, and Lockout Mode is
selected, the pump will enter Auto-Run Mode. When the pump is powered on, it will immediately begin
executing the Pumping Program. The user will immediately be prompted to enter a sub-program number.
This permits a set of production dispenses to be programmed into the pump. The user then would only be
able to select from one of these production dispenses and not make any changes.

Foot Switch Trigger Feature: If the Pumping Program was Paused, and the Paused Phase was a
continuous pumping Rate function, then starting the pump with a foot switch will cancel the Pause and
restart the Pumping Program Phase 1 causing the current program selection to be re-executed.

When the Program Selection Input function

is executed, the Pumping Program pauses and

displays:

nn

, where 'nn' is the Program Selection.

Using the right 2 arrow keys, under 'nn', the user enters the Label of the required Pumping Program. The
Pumping Program Label is defined by any number from 1 to 99. When the 'Start' key is pressed, the
Pumping Program continues execution at the Program Phase with the selected Pumping Program Selection
Label. Also, the accumulated dispense and withdrawal dispensed volumes are set to zero.

The pump searches for the selected Pumping Program Selection Label starting with the current Phase and
continuing to the end of the Pumping Program memory, then from Phase 1 until the current Phase is

reached again. If the selected label is not found, the 'out of range' error message

is displayed.

Pressing any key returns the display to the Program Selection Input display.

If more than one Phase is defined with the same label, then execution continues with the first matching
label encountered. The last selected program label is stored in non-volatile memory and becomes the
default label the next time the current Program Phase is executed. More than one Program Selection Input
function can be defined and placed at any Program Phase needed. To cancel the Program Select Input and
stop the Pumping Program, turn the power to the pump off and on.

If alarms are enabled, the buzzer will beep continuously while waiting for the start trigger.

11.3.7

'Pr:nn': Program Selection Label

The Sub-Program Start Label function

defines the start of a Pumping Program sub-program

that can be selected by the user during Pump Program execution.