New Era NE-9000 Series Programmable Peristaltic User Manual

Page 42

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-9000

Publication #1200-04

34

Phase

Function

1

OUT.0

Phase

Function

Rate

Volume

Direction

2

RATE

10.0 ml/min

0.005

Dispense

Phase

Function

3

EV:05

Phase

Function

Rate

Volume

Direction

4

RATE

10.0 ml/min

0.0 ml (off)

Dispense

Phase

Function

5

OUT.1

Phase

Function

Rate

Volume

Direction

6

RATE

10.0 ml/min

0.005

Dispense

Phase

Function

7

EV:01

Phase

Function

8

LP:ST

Phase

Function

Rate

Volume

Direction

9

INCR

1.0

0.25 ml

Dispense

Phase

Function

10

LP:14

Phase

Function

Rate

Volume

Direction

11

RATE

25.0 ml/min

0.0 ml (off)

Dispense

11.4.6

Example 6: Automated Dispensing with Synchronization

'Power Failure
Mode' Pumping
Program Restart
at Power Up

Syringe

Refill

Event Trigger
from
Syringe Filled Sensor:
Jump to Phase 3

5:00:00

5:00:00

5:00:00

5:00:00

5:00:00

Restart Pumping Program

The following is an automated dispensing Program for a NE-9000 peristaltic pump equipped with a
‘Syringe Filled Sensor’ attached to the TTL I/O connector, and a valve system to refill the syringe from a
reservoir. It is also assumed that the ‘Power Failure’ mode is enabled.

After a power fail restart, the pusher block is in an unknown position, making it impossible for an
automated dispensing system to regain synchronization. With the Syringe Filled Sensor, the following
Pumping Program will automatically synchronize the dispensing system, and then continue with the
normal dispense.

The first 2 Phases set an event trap for the Syringe Filled Sensor and refills the syringe until the sensor is
triggered. When the sensor triggers the event, the pump’s pusher block will be synchronized with the
Pumping Program. It is assumed that the sensor is positioned to refill the syringe with 60 ml. A withdraw
volume of 61 ml is set as a safety feature.