B. selection of filters – NTi Audio Minilyzer ML1 User Manual

Page 14

Advertising
background image

14

b.

Selection of Filters

Fig 2.3 Example of Filter Submenu

There are two sets of filters with the following characteristics

available:

Set 1:

FLAT

Flat frequency response (no filtering)

A-WTD A-weighting filter acc. IEC 60651

C-MESS C-Message filter acc. CCIR Rec. 468-4

HP22

Highpass 22 Hz acc. DIN 45045, -120 dB/dec.

HP60

Highpass 60 Hz acc. DIN 45045, -120 dB/dec.

HP400

Highpass 400 Hz acc. DIN 45045, -120 dB/dec.

VOICE

Voice band filter acc. ITU-T P.48

Set 1 filters are applicable with the measurement functions LEVEL

RMS, LEVEL REL, THD+N and SWEEP.

Set 2:

FLAT

Flat frequency response (no filtering)

A-WTD A-weighting filter acc. IEC 60651

C-WTD C-weighting filter acc. IEC 60651

X-CRV

-1

inverted X-Curve filter acc. ISO 2969

Set 2 filters are applicable with the measurement functions LEVEL-

SPL and 1/3

rd

OCT., whereby the X-CRV

-1

filter is enabled in the

1/3

rd

OCT. mode only.

Any selected filter will stay active until it is disabled by the user.

However, if the filter is not supported by the measurement function,

it will be disabled automatically.

Basic Operation

Advertising