4. INDUCTION LOOP MODE (OPTIONAL)

The induction loop mode is available upon installation of the MiniLINK

PC interface. The Minilyzer simplifies verifications and adjustments of

AFILS (Audio Frequency Induction Loop Systems). AFIL systems are

used in public buildings to increase the speech intelligibility for users of

hearing aids. Most hearing aids have a built in magnetic coil receiving

audio signals generated by induction loop systems. AFIL systems

consist of a wire loop and a current amplifier generating a magnetic

field into the hearing aid, modulated by the audio input signal, e.g.

voice. Adjustment and verification of AFIL systems is described in the

IEC 60118 standard. The Minilyzer, together with an induction

loop receiver, is an ideal tool to accomplish the necessary

measurements.

Induction Loop Receivers

There are various induction loop receivers available from different

AFILS components manufactures. All known types work together with

the ML1. Some receivers have a built in A-weighting filter, which must

be disabled to prevent measurement errors.

Activation of the Induction Loop Mode

Fig 4.1 Setup Induction Loop Mode

• Enter the setup page of the

ML1

• Change the position “Ind.

loop mode” to “ENABLE”

• Press ESC key to exit the

setup page

After this the ML1 restarts in the induction loop measurement mode. To

return to the common ML1 measurement mode, disable the induction

loop mode in the setup page again.