c. Calibration Panel

The Minilyzer ML1 enables the calibration of microphones, such as

the MiniSPL.

Fig 3.8 Calibration Panel

Select CALIB. in the LEVEL SPL panel field. Press the enter key and

the calibration screen SET SENSITIVITY ... will be displayed (Fig 14).

The following three calibration modes are available:

TO DEFAULT MINISPL-MIC: All MiniSPL are factory calibrated to a

sensitivity of 20.0 mV/Pa. Reset to default setting:
• Select the GO! field on the left.

• Press the enter key to start.

• OK! is shown to confirm the completed calibration.

USING EXT. REF.: The sensitivity may be adjusted with an external

calibrator to produce a certain calibrated sound pressure level.
• Select the dB

SPL

-value, press the enter key and adjust with the

cursor keys the sound pressure level generated by the calibrator.

Use up/down keys for 1.0 dB

SPL

steps and left/right keys for 0.1

dB

SPL

steps. The setting range 80 - 140 dB

SPL

is supported.

• Confirm the setting with the enter key.

• Produce the specific reference signal with the calibrator to the

microphone attached.

• Press the left cursor key to select the GO! field and press enter to

execute the calibration, whilst the reference signal of the calibrator

is still present.

• The frame WORKING ... followed by the frame CALIBRATION

FINISHED! is shown centered on the calibration screen if all has

been found in good order. Otherwise various error indications may