Octave spl – NTi Audio Minilyzer ML1 User Manual

Page 38

Advertising
background image

38

Measurement Functions

Zoom (Y-Axis): The scaling of the spectrum’s Y-axis (sensitivity) is

fixed and may be adjusted manually.
• Select the zoom (Y-axis) field and press the enter key.

• Use the up/down keys to scroll the displayed level along the Y-

axis and the left/right keys zoom out/in the Y-axis, e.g. alter the

resolution of the division.

Time Constant: The time constant corresponds to the integration time

of the 1/3

rd

octave recording. It may be set to five value: 0.2 / 0.5 /

1.0 / 2.0 / 5.0 seconds. A lower time constant results in a quicker,

more ‘nervous' display, whilst a higher time constant ‘averages’

more samples, thus providing a more stable spectrum.

Bargraph: Below the spectrum, the fullband input RMS level (20 Hz

- 20 kHz) is indicated by a bargraph without auto ranging. Select the

different indication ranges by moving the cursor to the left or right

arrow below RNGE and press enter. The available ranges are:
• -120 to -20 dBu, residual noise of ML1 < -120dBu (<1μV), enables

measurement of low signal levels

• -100 to 0 dBu

• -80 to +20 dBu, please note in this range the residual noise of

ML1 is gained by 20 dBu.

1/3

rd

Octave SPL

The 1/3

rd

OCTAVE SPL screen shows the signal frequency spectrum

of the acoustic signal, divided into 31 bands.

Fig 3.20 1/3

rd

Octave SPL Screen

Arrow

Mode

Arrow Readout

Bargraph of Level SPL

Scaling

of Y-Axis

Zoom

(Y-Axis)

Time

Constant

1/3rd Octave

Spectrum

Run/Pause