THd+N

In this mode the Minilyzer measures the THD+N (Total Harmonic

Distortion + Noise) and the k2 - k5 harmonic distortion.

Fig 3.10 THD+N Meter Panel

Input Level

(RMS)

The measurement is carried out in the band from 10 Hz - 20 kHz,

expressed in dB or in %. Simultaneously, the LCD shows the actual

RMS input level or the 2nd, 3rd, 4th & 5th harmonics distortion below

the THD+N result.

NOTE

The THD+N results are calculated using a restricted

measuring bandwidth of 10 Hz - 20 kHz.

Besides checking the linear purity of a sine signal - e.g. measuring

amount of harmonic distortions - this measurement mode is

particularly suited to get a quick idea whether unwanted disturbances

like hum are present.