NTi Audio Minilyzer ML1 User Manual

Page 36

Advertising
background image

36

A dialog box will open (Fig

3.18) on the graph, where the

sweep duration (min. 60 sec

- max. battery lifetime) and the

recording interval (≥ 1 sec) have

to be entered. Please note that

the number of samples must be

between 60 - 1600.

Measurement Functions

Fig 3.18 Time Sweep Setup

• To start the sweep recording, move the cursor to the GO! field

and press enter.

• The time sweep automatically stops when the defined duration

has expired. Alternatively, the sweep may be stopped manually

by pressing enter with the cursor on the REC field.

• To display the mean or the min-max values of the recorded sweep,

place the cursor to the corresponding field and select the required

mode by using the enter and the arrow keys.

• In order to analyze the sampled curve more detailed, activate the

arrow mode by placing the cursor to the corresponding symbol

and pressing the enter key. Move the readout line to the sample(s)

of interest by using the left / right keys, or press the up/down keys

to zoom out/in the X-axis.

• To zoom in/out the Y-axis, move the cursor to the zoom mode

field, press enter and

- use the up/down keys to zoom out/in the Y-axis

- use the left/right keys to scroll through the Y-axis.

The last time sweep recording will be stored internally, even after

leaving the sweep mode or switching off the Minilyzer. As soon as the

time sweep mode is re-entered, the last displayed curve will appear

on the graph, until a new time sweep recording is started.

NOTES • As soon as the sweep mode is entered, the filter

that has been active during the recording of the last

curve will be re-activated.

• If a running time sweep is stopped due to empty

batteries, the recorded data will not be lost.

• The auto power off is disabled during a time

sweep.

Advertising