Signal balance error – NTi Audio Minilyzer ML1 User Manual

Page 32

Advertising
background image

32

Measurement Functions

signal balance error

The signal balance error reflects the deviation from the perfect

balance status. No signal balance error indicates that the absolute

levels at XLR-pin 2 and pin 3 relative to pin 1 (ground) are identical

and opposite in polarity.

Fig 3.15 Signal Balance Error Screen

Frequency of

Input Signal

Balance Indicator

XLR-pin 2,3

Signal Balance

Error

Input

Level-RMS

Bargraph

Zoom

Control

Bargraph

Zoom Mode

Bargraph with Scaling Information

In the BALANCE mode, the signal balance error is shown in a numerical

value, which is the deviation from the optimum in percent.
balance Indicator: The direction of the deviation is indicated by the

arrows like 2 < 3 or 2 > 3, whereby the numbers 2 and 3 represent

the signal at XLR-pin 2 and pin 3.
In the Level, THD+N and Polarity (electrical input only) measurement

function, the Minilyzer permanently monitors the balance of the input

signal by a graphical indicator. This enables e.g. to check whether

cable connections are made correctly.

Bargraph: The bargraph shows an analog display of the signal

balance error. The scaling may be controlled automatically or

manually.
• Select manual (M) or automatic (A) scaling by the bargraph zoom

mode field.

• Within the manual scaling press the left/right keys to scroll through

the actual range or the up/down keys to increase or decrease the

range (sensitivity) of the bargraph scale.

• Press enter to confirm your setting.