Calibration

An induction loop receiver transforms a magnetic field strength to an

electrical level. The receiver sensitivity is specified within the technical

data. Before a valid measurement can be started the ML1 sensitivity

setting must be adjusted accordingly.

Fig 4.3 Calibration Panel

filters

The ML1 has a built-in A-weighting and a HP400 (high pass 400Hz)

filter, which may be activated using the filter menu. For detailed test

requirements please refer to the IEC 60118 standard. The HP400 filter

has an excellent rejection of mains frequencies, which sometimes is a

dominant part of the signal received by the induction loop receiver.

Fig 4.4 SInduction Loop Filters

Induction Loop Mode