Vu + ppm – NTi Audio Minilyzer ML1 User Manual

Page 28

Advertising
background image

28

vu + PPM

ML1 provides vu + PPM (Peak Program Meter) measurement results,

following IEC60268.

Reference

Level

Numerical

Peak-Hold
Graphical

Peak-Hold

Reference Indicator

Integration Time

Fig 3.11 vu + PPM (Type I) Panel

The PPM may be measured according to three standards:

Type I

Type IIA

Nordic standard

One of these standards has to be selected when entering the vu+PPM

mode through the measurement functions menu.

The following chapters describe the different entry fields and indicators

of the vu+PPM display.

a.

Reference Level & Indicator

The reference level indicates the RMS level of a 1 kHz sine wave that

defines the value of the reference indicator. The reference level may

be defined by the user in a range of -20 dBu to +14 dBu.

The default reference level depends on the selected standard

+4 dBu for the vu-meter

+6 dBu for the PPM standards Type I and Nordic

+8 dBu for the PPM standard Type IIA.