Time weighting: All sound pressure level measurements include

time weighting. A change in sound pressure level is read out on

the display with a shorter or longer response time.
Available response times, corresponding to IEC 60651, are
• SLOW (long attack- and release time)

• FAST (short attack- and release time)

• IMP (=impulse, very short attack and long release time)

Bargraph: The bargraph provides an analog display of the actual

sound pressure level.
Available bargraph ranges are:
• 20 - 100 dB

SPL

• 40 - 120 dB

SPL

• 60 - 140 dB

SPL

The selected bargraph range complies to the measurement range

of the instrument. To achieve accuracy of the measurement within

the specified tolerances the proper range must be selected, called

primary indicator range. The two range indicator arrows assist to

set the Minilyzer to the optimal range. As soon as the indicated

bargraph value is found lower than the primary indicator range, the

down arrow symbol below RNGE will start to move, symbolizing the

actual bargraph reading is inaccurate. The lower bargraph range

needs to be changed manually by selecting the moving left arrow

symbol and pressing the enter key.

Fig 3.4 SPL Panel with Overload