Sound Pressure Level

In the LEVEL SPL mode the Minilyzer measures the integrated-

averaging level over time, the actual, minimum and maximum sound

pressure level.

NOTES • For this function a self powered microphone is

required. NTi Audio does recommend to use the

accessory MiniSPL (for details see Accessories).

• Prior the first measurement the Minilyzer has to

be calibrated to the microphone specifications.

As default the ML1 is calibrated according the

Minispl.

• No phantom power is available on the XLR-input.

a. Sound Pressure Level Panel

Fig 3.3 Sound Pressure Level Panel

Calibration

Minimum,

Maximum or

Actual SPL

Pause

Clock

Equivalent

Continious

Sound

Pressure

Level

Time

Weighting

SPL Bargraph with manual ranging

To enter the LEVEL SPL mode (Fig 3.3), select LEVEL -> SPL in the

measurement functions submenu.
For most of the ML1 applications, the sound pressure level

measurements require the use of the A-weighting filter. To prevent

wrong filter settings, causing wrong non-usable test results, the default

filter setting of the “LEVEL->SPL” mode is A-weighting (A-WTD).

Note:

Each time the LEVEL SPL mode is selected the

A-weighting filter is pre-set by default.