Measurement resolution – Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual
Page 73

Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012
73
Setting Up Measurements
Chapter 2
Measurement Resolution
The measurement resolution is part of the collection specification (Setup >
Collection).
See Setting Up Collection Specifications on page 69
. You select the
collection specification when you set up the measurement definition.
Emonitor and the Enpac 2500 support the following resolutions.
The values that appear in the Number of lines drop down list in the Collection
Specification dialog are those for spectrum measurements. Emonitor uses the
corresponding number of time samples from the table above for time
measurement definition.
You can calculate the number of time samples from the number of spectral lines
using the following formula.
You can calculate the total time required to collect a time waveform measurement
from the Number of lines value and the Maximum frequency value in the
Collection Specification dialog (Setup > Collections).
If your Maximum frequency is specified in CPM, convert CPM to Hertz using
the following formula.
Table 16 - Measurement Resolutions
Spectral Lines
Time Samples
100
256
200
512
400
1024
800
2048
1600
4096
3200
8192
6400
16,384
12,800
32,768
time samples
spectral lines 2.56
=
Fmax (Hz)
Fmax (CPM)
60
-------------------------------
=