Enwatch Unit 19

Bias voltage reading

When you define a process measurement of DC in the Emonitor software, the Enwatch unit reads the transducer bias
voltage.
24V means open circuit, 10V is OK.

Maximum number of averages that can be supported

Number of samples required: Lines x 2.56 + lines x 2.56(#avg -1) x (1 -%overlap)
Max number of samples per configuration: 32768
Max number of averages = ((32768 / (lines x 2.56)) - 1) / (1 -%overlap) + 1)

The following table shows the maximum number of averages versus the number of lines and percent overlap for non-trigger

point.

The following table shows the maximum number of averages versus the number of lines and percent overlap for pre-trigger

point.

Signal detection

RMS, Peak, Peak-Peak

Configuration downloading

Route mode configuration downloads to Flash memory
Live mode configuration downloads to RAM

Note: You can reduce number of average or increase percent overlap to reduce amount of memory used.

Table 2 - Enwatch measurement capabilities

Product Feature

Capability

Max samples-non trigger points 32768

% overlapping

% overlapping

% overlapping

% overlapping

#lines

0

0.25

0.5

0.75

100

128

170 255 509

200

64 85 127

253

400

32

42 63 125

800

16

21 31 61

1600

8

10

15

29

3200

4

5 7 13

6400

2

2 3 5

12800

1

1 1 1

Max samples-trigger points 16384

% overlapping

% overlapping

% overlapping

% overlapping

#lines

0

0.25

0.5

0.75

100

64 85 127 253

200

32 42 63 125

400

16 21 31

61

800

8 10 15 29

1600

4 5 7 13

3200

2 2 3 5

6400

1 1 1 1

12800

-

-

-

-