Figure 50, Msm – components – Metrohm 850 Professional IC Anion MCS LP-Gradient User Manual

Page 108

Advertising
background image

5.11 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

98

■■■■■■■■

850 Professional IC – Anion – MCS – LP Gradient

1

2

3

4

5

Figure 50

MSM – Components

1

Union nut

2

MSM connecting piece 6.2832.010

3

MSM rotor A 6.2832.000

4

MSM housing

5

Slot in the MSM housing

Cleaning the MSM

Clean the MSM as follows:

1 Disconnecting the MSM from the IC system

Switch off the instrument.

Disconnect the MSM from the separation column, peristaltic
pump and detector.

2 Dismantling the MSM

Unscrew union nut (50-1) from the MSM housing (50-4).

Pull MSM connecting piece (50-2) and MSM rotor A (50-3) out of
the MSM housing (50-4). The MSM connecting piece and MSM
rotor A normally stick to one another - if this is not the case: Take
a sharp object, insert into the slot (50-5) in the MSM housing,
and pull out MSM rotor A (50-3) in this way.

Detach the MSM connecting piece (50-2) from MSM rotor A
(50-3).

3 Cleaning the supply and discharge lines

Connect in turn each of the 6 capillary tubings fastened on the
MSM connecting piece (50-2) on the high pressure pump (see
Chapter 3.12, page 42)
and pump through ultra pure water.