8 high pressure pump, High pressure pump – Metrohm 850 Professional IC Anion MCS LP-Gradient User Manual

Page 125

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Technical specifications

850 Professional IC – Anion – MCS – LP Gradient

■■■■■■■■

115

7.8

High pressure pump

Type

Serial dual-piston pump

Intelligent pump head recognition

Chemically inert

Metal-free pump heads

Materials in contact with eluent: PEEK, ZrO

2

, PTFE/PE

Self-optimizing flow and pressure

Flow rate

Adjustable flow
range

0.001…20.0 mL/min

Flow increment

1 µL/min with standard pump head

Reproducibility
of the eluent
flow

< 0.1 % deviation

Pressure range

Pump

0…50.0 MPa (0…500 bar)

Pump head

0…35.0 MPa (0…350 bar) (applies for the standard PEEK pump head)

Residual pulsa-
tion

< 1 %

Safety shutdown

Function

Automatic shutdown upon reaching the pressure limit values

Maximum pres-
sure limit

Adjustable from 0.1…50 MPa (1…500 bar)

The pump is automatically shut down at the first piston stroke
above the maximum limit value

Minimum pres-
sure limit

Adjustable from 0…49 MPa (0…490 bar)

The shutdown mechanism is inactive at 0 MPa

The shutdown mechanism only becomes active 2 minutes after sys-
tem start

The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit

Advertising