6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

112

■■■■■■■■

850 Professional IC – Anion – MCS – LP Gradient

Problem

Cause

Remedy

Eluent – Gas bubbles in the
eluent.

Check connections of the eluent degasser (see
Chapter 3.10, page 37)
.

High pressure pump –
defective.

Request Metrohm Service.

Vacuum is not being
built

Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.

Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module:

Connect an FEP tubing between the IC
instrument and the Extension Module and
tighten both clamping screws until they
seal.