8 capillary and cable feed-throughs, Capillary and cable feed-throughs – Metrohm 850 Professional IC Anion MCS LP-Gradient User Manual

Page 40

Advertising
background image

3.8 Capillary and cable feed-throughs

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

850 Professional IC – Anion – MCS – LP Gradient

7

Drainage tubing
Section of the 6.1816.020 silicon tubing.
Guides escaped fluid into a waste container.

8

Drainage tubing connection
For draining escaped fluid from the base tray
through the connected drainage tubing.

9

Drainage tubing connection
For supplying escaped fluid through the con-
nected drainage tubing to the leak sensor.

Proceed as follows to install the drainage tubing:

Installing drainage tubing

1 Plug drainage tubing (9-2) into the drainage tubing connection (9-1)

of the covering plate and shorten to required length.

2 Plug drainage tubing (9-4) into the drainage tubing connection (9-3)

of the detector chamber and shorten to required length.

3 Connect drainage tubing (9-2) from the covering plate and drainage

tubing (9-4) from the detector chamber to the Y connector (9-5).

4 Connect drainage tubing (9-6) to the Y connector (9-5), shorten to

required length and plug the other end of the drainage tubing into
the drainage tubing connection (9-9).

5 Plug drainage tubing (9-7) into the drainage tubing connection (9-8)

of the base tray and guide the other end into a waste container.

3.8

Capillary and cable feed-throughs

Several openings have been integrated for feeding through capillaries and
cables. They are located on the doors (see Figure 10, page 31), on the
rear panel (see Figure 7, Page 25) or below the covering plate or above
the base tray (see Figure 11, Page 32).