3 Installation

850 Professional IC – Anion – MCS – LP Gradient

■■■■■■■■

41

Connecting connection tubings

Note

The connection tubings to the inputs B and C are already installed at
the time the instrument is delivered. Only connector A still needs to be
connected.

4

5

1

2

4

5

4

5

3

6

Figure 20

Connecting eluent connection tubings

1

Connection tubing eluent degasser –
mixing valve (6.1834.120)

2

Connection tubing eluent degasser –
mixing valve (6.1834.100)
Pre-installed.

3

Connection tubing eluent degasser –
mixing valve (6.1834.110)
Pre-installed.

4

Backup ring

5

Clamping screw

6

Mixing coil for low pressure gradients
6.2758.020

1

Plug the connection tubing (20-1) into eluent inlet A and fix in
place with the backup ring (20-4).

Carefully tighten the clamping screw (20-5) with 6.2621.050
wrench.

2 Repeat Step 1 with the other two eluent connection tubings (20-2)

and (20-3).