3.19 Peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

60

■■■■■■■■

850 Professional IC – Anion – MCS – LP Gradient

3 Connecting the pressure side

Note

Depending on the use of the peristaltic pump, on the pressure side
you can either connect:

Case A: a 6.2744.180 pump tubing connection with filter
(see Figure 33, page 60) or

Case B: a 6.2744.160 pump tubing connection without filter
(see Figure 34, page 61).

For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection with filter must be used.

Case A: 6.2744.180 pump tubing connection with filter:

1

2

3

Figure 33

Install pump tubing connection with filter

1

Union nut

2

Adapter

3

Tubing olive with filter holder

Slide union nut (33-1) onto the pump tubing.

Select a suitable adapter (33-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 2,
page 61)
.

Place the tubing olive with filter holder (33-3) onto the pump tub-
ing.

Screw the union nut (33-1) onto the tubing olive (33-3).

or

Case B: 6.2744.160 pump tubing connection without filter: