3 Installation

850 Professional IC – Anion – MCS – LP Gradient

■■■■■■■■

39

Table 1

Smallest possible eluent percentage depending on installed
pump head and set flow

Flow

Standard pump head

Macro pump head

1 mL/min

10 %

2.5 %

2 mL/min

20 %

5 %

3 mL/min

not recommended

7.5 %

Depending on the application, an optimal assignment of each eluent to a
valve is documented in the Application Notes. To obtain a reproducible
gradient profile, this eluent-valve assignment must remain unchanged.

Each eluent is degassed by means of its own eluent degasser (see Chapter
3.10, page 37)
.

1

2

3

4

6

5

7

8

9

Figure 18

Low pressure gradient

1

Mixing valve

2

Eluent degasser
for Eluent B

3

Eluent degasser
for Eluent C

4

Connection tubing 6.1834.100
Connects eluent degasser (18-

3

) and mixing

valve C. Pre-installed.

5

Eluent inlet C

6

Connection tubing 6.1834.100
Connects the eluent degasser (18-

2

) and the

mixing valve input B. Pre-installed.

7

Eluent inlet B

8

Eluent inlet A

9

Eluent outlet