3.18 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

56

■■■■■■■■

850 Professional IC – Anion – MCS – LP Gradient

1

3

2

3

4

5

6

7

8

2

1

Figure 30

MSM – connections

1

Union nut

2

MSM connecting piece 6.2832.010

3

Eluent input capillary
Labeled with Eluent.

4

Eluent output capillary
Labeled with Detector.

5

Rinsing solution input capillary
Labeled with H2O.

6

Rinsing solution output capillary
Labeled with Waste.

7

Regeneration solution output capillary
Labeled with Waste.

8

Regeneration solution input capillary
Labeled with H2SO4.

The PTFE capillaries firmly mounted on the MSM are connected to the
other components of the IC system as follows:

Caution

As the PTFE capillaries are very soft, the pressure screws should not be
overtightened.

Flattened capillaries must be shortened with the help of a 6.2621.080
capillary cutter.