Figure 19, Connecting mixing coil for low pressure gradients – Metrohm 850 Professional IC Anion MCS LP-Gradient User Manual

Page 50

Advertising
background image

3.11 Low pressure gradient

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

850 Professional IC – Anion – MCS – LP Gradient

Connecting mixing coil for low pressure gradients

2

3

4

1

Figure 19

Connecting mixing coil for low pressure gradients

1

Eluent outlet

2

Pressure screw long
for the connector on the mixing valve.

3

Mixing coil for low pressure gradients
6.2758.020

4

Pressure screw short
for the connection on the high pressure
pump.

1 Screw on the 6.2758.020 mixing coil for low pressure gradients

(19-3) with the 6.2744.090 long PEEK pressure screw (19-2) to the
eluent outlet (19-1).

2 Fasten the other end of the mixing coil for low pressure gradients

(19-3) with the 6.2744.070 short PEEK pressure screw (19-4) directly
to the inlet valve holder of the high pressure pump (see Chapter
3.12.2, page 43)
.