3.2. Remote control commands

794 Basic Titrino

106

Mode.Parameter.Statistics.ResTab.Selected original, delete n,

delete

all

Mode.Parameter.Statistics.ResTab.DelN

1...20

Entries for the statistics calculations.

.Status:

On/off switching. Requirement for statistics calculations is a valid

assignment, see 3.2.2.9.

.MeanN: Number of individual results for statistics calculations.

.ResTab.Select: Selection of the table for the statistics calculations.

original: Original table. The original table is (again) set up, i.e. any

individual results which have been deleted are reincorporated

in the statistics calculations.

delete n: Single result lines are removed from the statistics calcu-

lation. All results of the corresponding line in the statistics

table are deleted. Specification of the line number in

.ResTab.DelN.

delete all: Clear entire statistics table. The results can not be reac-

tivated.

.ResTab.DelN: Specification of the line number to be deleted.

3.2.2.24. Mode.Parameter.Evaluation.EPC DET:

0...5...200

MET

pH:

0.1...0.50...9.99

U,

Ipol:

1...30...999

Upol:

0.1...2...99.9

Mode.Parameter.Evaluation.Recognition.Selected

all, greatest,

last, window, OFF

Mode.Parameter.Evaluation.Recognition.Window.1.LowLim

pH:

0...±20.00, OFF

U,

Ipol:

0...±2000, OFF

Upol:

0...±200.0, OFF

Mode.Parameter.Evaluation.Recognition.Window.1.UpLim

Input range as LowLim

etc. up to 9 windows

Mode.Parameter.Evaluation.FixEP.1.Value pH:

0...±20.00, OFF

U;

Ipol:

0...±2000, OFF

etc. up to 9 fix EP's

Upol: 0...±200.0, OFF

Mode.Parameter.Evaluation.pK ON,

OFF

Parameters for DET and MET: Evaluation of the EP's, see page 32.

.EPC: EP criterion in pH, in mV (with U and Ipol) or in µA (with Upol).

.Recognition.Selected: EP recognition.

all: All endpoints found are recognized.

great: Only the largest EP is recognized.

last: Only the last EP is recognized.

window: Only EP's that lie within set windows are recognized.

OFF: The EP evaluation is switched off.

.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U

and Ipol) or µA (with Upol).

.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and

Ipol) or µA (with Upol).

Windows are opened until the lower limit is set to OFF. For every

expected EP, an individual window must be set, see page 34.

.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. µA (for Upol). Fix EP's are

evaluated until the setting OFF is found.