2.7 Parameters, key <PARAM>
794 Basic Titrino
41
Titration sequence of SET
After the start, the activate pulse is outputted.
The start delay time is waited off.
If conditioning is on, the sample solution is titrated until
the (first) EP is reached. The display shows then
drift OK 2.3 µl/min
or
SET pH conditioning
The vessel is now conditioned. The titration can be started
with <START>.
The sample identifications and the sample size are re-
quested.
The temperature is measured if a T sensor is connected.
The
start volume is dispensed and pause 2 waited off.
Then the titration is performed to the first, then to the sec-
ond end point.
Calculations are carried out.
Data are outputted.
Conditioning is carried out.
(Activate pulse)
<START>
(Start delay)
(Start conditions)
(Request ident.)
(Request smpl size)
Titration:
1
st
end point
2
nd
end point
Calculations
Data output
(Reconditioning)
(Preconditioning)
(<START>
(Activate pulse)
(Start delay)