6.3 User methods
794 Basic Titrino
179
FormolNo TIP sequence
-
Start Form.Pre
-
Activation of Dosimat
-
Reaction time
-
Start Form.Det
-
Formaldehyde number in mL c(NaOH) = 0.1
mol/L per 100 mL sample
-
Calculation factor for 100 mL sample
-
EP1 from Form.Det
'pa
794 Titrino 01102 794.0010
date 2002-01-03 time 10:00 0
TIP FormolNo
parameters
>sequence
1.method: Form.Pre
2.L6 output: pulse
3.pause 60 s
4.method: Form.Det
>statistics
status: ON
mean n= 3
res.tab: original
>preselections
req.ident: OFF
req.smpl size: OFF
meas.mode: OFF
temperature 25.0 °C
------------
'fm
794 Titrino 01102 794.0010
date 2002-01-03 time 10:00 0
TIP FormolNo
>calculations
FormolNo=C70*C01;1;
C01= 4
C70= 5.522
------------