User methods – Metrohm 794 Basic Titrino User Manual
Page 182

6.3. User methods
794 Basic Titrino
178
Form.Det Determination of the sample.
- Transfer of the end point as temporary variable
to FormolNo.
'pa
794 Titrino 01102 794.0010
date 2002-01-03 time 09:58 0
SET pH Form.Det
parameters
>SET1
EP at pH 8.50
dynamics 1.5
max.rate 10.0 ml/min
min.rate 25.0 µl/min
stop crit: drift
stop drift 20 µl/min
>SET2
EP at pH OFF
>titration parameters
titr.direction: +
start V: OFF
pause 0 s
meas.input: 1
temperature 25.0 °C
>stop conditions
stop V: abs.
stop V 99.99 ml
filling rate max. ml/min
>statistics
status: OFF
>preselections
conditioning: OFF
req.ident: OFF
req.smpl size: OFF
activate pulse: OFF
------------
'de
794 Titrino 01102 794.0010
date 2002-01-03 time 09:59
SET pH Form.Det
def
>formula
>silo calculations
match id: OFF
>common variables
>report
>mean
MN1=RS1
>temporary variables
C70=EP1
------------