6.1 Technical specifications
794 Basic Titrino
159
Display
LCD, 2 lines of 24 characters each
Height of characters 5 mm
LED
back-lit
Internal memory
Method memory for up to 100 methods.
Data bank with 17 Metrohm methods.
Silo memory for sample data and results
RS232 interface
for printer, balance or computer connection: completely
controllable from external control unit
Remote input/output lines for Sample Changer, robot connection, etc.
Analog output
Output signal
-2000 ... 2000 mV
Signal at analog-
depending on preselection:
output
U (measuring value)
dU/dt (measured value drift)
V (volume)
dV/dt (volume drift)
U(rel) (control deviation at SET)
T (temperature)
1 mV (12 Bit), see also page 153
Ambient temperature
Nom. operation range
5 ... 40 °C
Storage
– 20 ... 60 °C
Transport
– 40 ... 60 °C
Safety specifications
Designed and tested in accordance to IEC publication
61010-1, safety class I. This manual contains information
and warnings which have to be followed by the user to en-
sure safe operation and to retain the apparatus in safe
condition.