2.7. Parameters, key <PARAM>

794 Basic Titrino

48






activate pulse: OFF

Calibration temperature and pH values of the buffers

(which are temperature dependent) must therefore be

entered in advance. The inputs in key <PARAM> are

valid.

Pulse output on the line "activate" (L6, pin 1) of the

remote socket (all, first, OFF)

See page 163.

>statistics

status: OFF




mean n= 2

res.tab: original






delete n= 1

Statistics calculation

Status of statistics calculation (OFF, ON)

If the statistics calculation is switched off, the following

inquiries regarding the statistics do not appear.

Mean value calculation from n single results (2...20)

Result table (original, delete n, delete all)

"original":

The original table is used. Deleted

individual results are again incorporated in

the evaluation.

"delete n": Deletion of single results with the index n.

"delete all": The entire table is deleted.

Delete data from sample number n (1...20)

The deleted result is removed from the statistics

calculation.