2.4. Tutorial

794 Basic Titrino

14

If you have a printer connected, you probably wish to have the curve and a result report

printed out automatically at the end of a titration.

To install a printer, see page 150.

Print reports

4 x <DEF>

<ENTER>

6 x <SELECT>

<;>

9 x <SELECT>

<ENTER>

<QUIT>

• Press <DEF> repeatedly until the display shows:

• Press <ENTER> to move to the definition of reports.

• With <SELECT> you select the individual repot

blocks. Use a ";" as delimiter between the report

blocks to print more than one block. If you wish to

print out a curve and a full result report, enter

"curve;full".

• Confirm the entry with <ENTER> and quit the inquir-

ies with <QUIT>.

Now print your reports by pressing <PRINT><REPORTS><ENTER>. Your printout will

look like the following:

Identification of the report type (cu=curve)

Start volume, mode and method name

Scaling of curve axis

Identification of the report type (fr=full report)

Device type with identification and program version

Initial pH, mode and method name

Volume and pH value of EP1

Calculated result

def

>report

report:

report:curve;full

'cu

794 Titrino 01102 794.0010

date 2002-01-03 time 17:34 1

start V 0.000 ml DET pH ********

1.0 ml/div dpH=2.0/div


−−−−−−−−−−−−−−−−

'fr

794 Titrino 01102 794.0010

date 2002-01-03 time 17:34 1

pH(init) 2.88 DET pH ********

smpl size 2 ml

EP1 2.083 ml 7.64

RS1 3.80 g/l

manual stop
−−−−−−−−−−−−−−−−