Metrohm 794 Basic Titrino User Manual

Page 16

Advertising
background image

2.4. Tutorial

794 Basic Titrino

12

If more than one equivalence point has been found, the others can be viewed with

<ENTER>.

Calculation of the result: formula entry

<DEF>

<ENTER>

<1>

<ENTER>

<SELECT>

<ENTER>

2 x <QUIT>

With this equivalence point a result can be calculated.
• Press the <DEF> key.


• Press <ENTER> to move on to the formula entry.

The display now shows "RS?".

• Press "1", i.e. the first formula.

You now can enter a formula. Note here the top inscrip-

tion on the keys of the keypad and the numbers. The

following symbols can be used:
EP# EP's with 1-digit number, e.g. EP1.

RS# Previously calculated results, e.g. RS1 in the sec-

ond formula.

C## Calculation constant with 2-digit number, e.g. C01.

C00 is reserved for the sample size. For the mean-

ing of the calculation constants see page 53.

Calculate the content of your hydrochloric acid in g/L:

RS1=EP1*C01*C02/C00

End point*conc.(titrant)*molecular weight/sample size

To correct a formula, delete it with <CLEAR>.
• Confirm the formula with <ENTER>.

• You may enter a text for the result output, see page

8.

• Enter the desired number of decimal places for the

result.

• Select the desired unit g/L with <SELECT>, or type

a text as unit, see page 8.

The Titrino prompts for the calculation of the next result.

Quit the formula entry by pressing <QUIT> twice.

In order to be able to calculate the result, enter the calculation constants used in the for-

mula.

def

>formula

RS?

RS1=

RS1=EP1*C01*C02/C00

>formula

RS1 text

RS1

>formula

RS1 decimal places

2

>formula

RS1 unit

%

RS?

Advertising