Metrohm 794 Basic Titrino User Manual
Page 17

2.4
Tutorial
794 Basic Titrino
13
Entry of the calculation constants
<C-FMLA>
• Press <C-FMLA>.
The constants which have been used in the formula
are requested: input with digit keys, confirm with
<ENTER>.
C01: Concentration of your titrant= 0.1 mol/L. Enter
0.1.
C02: Molar mass of HCl = 36.47 g/mol
Entry of the sample size
4 x <SMPL DATA>
<2>
<ENTER>
<SELECT>
<ENTER>
• Press <SMPL DATA> repeatedly until "sample size"
appears in the display.
• Enter 2.
• Use <SELECT> to select the unit "mL" and confirm
the new value with <ENTER>.
The result is now calculated and can be displayed in place of the equivalence point. If
your method already includes a formula at the end of the titration, the calculated result is
displayed directly after the titration. As we have entered the formula later, we now have to
select the result display:
Display of the result
<SELECT>
<ENTER>
• Press <SELECT> repeatedly until ">display results"
appears in the display.
• Press <ENTER> to move to the result display.
C-fmla
>C01 0.0
smpl data
smpl size
1.0 g
smpl data
smpl unit:
g
DET pH
********
>display results
DET pH
********
RS1 3.80
g/l