3.2. Remote control commands

794 Basic Titrino

122

The cycle number is set to 0 on switching on the instrument and it

is always incremented as long as the instrument remains

switched on.

.V: Volume

.Meas:

Measured value associated to the cycle number.

The unit "assembly" must be preset (see 3.2.2.94).

3.2.2.96. Setup.SendMeas.Titrator.CyclNo ON, OFF

Setup.SendMeas.Titrator.V ON, OFF

Setup.SendMeas.Titrator.Meas ON, OFF

Setup.SendMeas.Titrator.dVdt ON, OFF

Setup.SendMeas.Titrator.dMeasdt ON, OFF

Setup.SendMeas.Titrator.dMeasdV ON, OFF

Setup.SendMeas.Titrator.ERC ON, OFF

Selection of the values from the titrator which are sent in the set time interval

(see 3.2.2.93, formats see 3.2.2.79):

.CyclNo: Cycle number. Together with the cycle time (3.2.2.82), a time

frame can be set up. The other data belong to the corresponding

cycle number. The cycle number is set to 0 at the start of a

method and it is incremented until the end of the method.

.V: Volume.

.Meas: Measuring

value

.dVdt:

associated volume drift.

.dMeasdt: associated measured value drift.

.dMeasdV: associated 1st derivative of the titration curve.

.ERC:

ERC in DET.

The unit "titrator" must be preset (see 3.2.2.94).

3.2.2.97. Setup.AutoInfo.Status ON, OFF

Setup.AutoInfo.P ON, OFF

Setup.AutoInfo.T.R ON, OFF

Setup.AutoInfo.T.G ON, OFF

Setup.AutoInfo.T.GC ON, OFF

Setup.AutoInfo.T.S ON, OFF

Setup.AutoInfo.T.B ON, OFF

Setup.AutoInfo.T.F ON, OFF

Setup.AutoInfo.T.E ON, OFF

Setup.AutoInfo.T.H ON, OFF

Setup.AutoInfo.T.C ON, OFF

Setup.AutoInfo.T.O ON, OFF

Setup.AutoInfo.T.N ON, OFF

Setup.AutoInfo.T.Re ON, OFF

Setup.AutoInfo.T.Si ON, OFF

Setup.AutoInfo.T.M ON, OFF

Setup.AutoInfo.T.EP ON, OFF

Setup.AutoInfo.T.RC ON, OFF

Setup.AutoInfo.I ON, OFF

Setup.AutoInfo.O ON, OFF

ON means that the Titrino reports automatically the moment the correspond-

ing change occurs.

.Status:

Global switch for all set AutoInfo.