2.4. Tutorial

794 Basic Titrino

20

After the second titration you receive a printout with statistics calculations:

If you have no printer connected, you can view the mean value and standard deviation:
• Press <SELECT> until "display mean" appears in the display.
• With <ENTER> you obtain the mean value.
• A second <ENTER> displays the number of single values which have been used for

the mean calculation.

• If you press <SELECT> again, the display "display std.deviation" appears. Once

again, you can view this value, by pressing <ENTER>.

Perhaps you have noticed that the two values differ too greatly? In any case, we shall per-

form a third determination with the same sample. The result of this determination will be

another value which is incorporated in the statistics calculations.

Addition of more determinations for the statistics calculations

5 × <PARAM>

<ENTER>

<ENTER>

<3>

<ENTER>

2 × <QUIT>

• Press <PARAM> until the display shows:


• Statistics calculations can be switched on either with

the <STATISTICS> key or in the inquiry "status:".

We shall leave them switched on and proceed to the

next inquiry.

Mean value of 2 single determinations.
• Enter "3" to include an additional determination.


• Quit the inquiry.

Perform another titration.

parameters

>statistics

>statistics

status: ON

>statistics

mean:

n= 2

'fr

794 Titrino 01102 794.0010

date 2002-01-03 time 17:34 3

pHc(init) 6.29 SET pH ********

EP1 0.0628 ml 4.26

m value 5.02 mmol/l

mean(2) +/s s/%

m value 5.04 0.028 mmol/l 0.56

==============