4 coefficient k(4) – Metrohm viva 1.0 Manual User Manual

Page 615

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

603

Unit

Concentration unit = #g/L

#A L

² g⁻² (for Evaluation quantity

= Height)

#C L

² g⁻² (for evaluation quantity

= Area)

Concentration unit = #L/L

#A (for Evaluation quantity =
Height)

#C (for Evaluation quantity =
Area)

Concentration unit = #mol/L

#A L

² mol⁻² (for Evaluation quan-

tity = Height)

#C L

² mol⁻² (for Evaluation quan-

tity = Area)

Remarks

The y axis of the calibration curve is dimensionless because of the fact that it
is standardized in the Dilution Titration (DT) calibration method. The effective
addition volume VEFF is always applied to the x axis. This means that the unit
for the coefficient k(2) = L

⁻².

5.6.7.4.4

Coefficient k(4)

Variable name

RS.’Command name’.’Substance name’.CALC4

Description

Fourth-order coefficient of the calibration curve or standard addition curve.

Decimal places

0…3…6

Unit

Concentration unit = #g/L

#A L

g⁻⁴ (for Evaluation quantity

= Height)

#C L

g⁻⁴ (for Evaluation quantity

= Area)

Concentration unit = #L/L

#A (for Evaluation quantity =
Height)

#C (for Evaluation quantity =
Area)

Concentration unit = #mol/L

#A L

mol⁻⁴ (for Evaluation quan-

tity = Height)

#C L

mol⁻⁴ (for Evaluation quan-

tity = Area)

Remarks

The y axis of the calibration curve is dimensionless because of the fact that it
is standardized in the Dilution Titration (DT) calibration method. The effective
addition volume VEFF is always applied to the x axis. This means that the unit
for the coefficient k(4) = L

⁻⁴.

Advertising