6.6 Dosing units subwindow

■■■■■■■■■■■■■■■■■■■■■■

676

■■■■■■■■

viva 1.0

Dosing rate Dosing port 1

Rate at which dosing should take place via Dosing port 1. The maximum
dosing rate depends on the cylinder volume of the dosing unit used (see
below). When the function is carried out the rate will automatically be
decreased to the highest possible value.

Input range

0.01 - 166.00 mL/min

Default value

4 mL/min

Selection

maximum

Dosing rate Dosing port 2

Rate at which dosing should take place via Dosing port 2. The maximum
dosing rate depends on the cylinder volume of the dosing unit used (see
below). When the function is carried out, the rate will automatically be
decreased to the highest possible value.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Dosing rate Fill port

Rate at which dosing or filling should take place via Fill port. The maxi-
mum dosing rate depends on the cylinder volume of the dosing unit used
(see below). When the function is carried out, the rate will automatically
be decreased to the highest possible value.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Dosing rate Special port

Rate at which dosing should take place via the Special port. The maxi-
mum dosing rate depends on the cylinder volume of the dosing unit used
(see below). When the function is carried out, the rate will automatically
be decreased to the highest possible value.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

Maximum dosing rate for the dosing unit in relation to the cylin-
der volume:

Cylinder volume

Maximum rate

2 mL

6.0 mL/min

5 mL

16.0 mL/min