ADS Environmental Services FlowShark QR 775002 A4 User Manual

Page 162

Advertising
background image

6-36

ADS FlowShark Manual

 Select the Close button to exit the Pressure 1 device dialog

and return to the Monitor Diagnostics dialog.

6. Select Activate from the Functions drop-down list, and then

select the Perform button. For more information on activating
a monitor, refer to Activating the Monitor on page 6-25.

7. Select the Disconnect button to discontinue communication

with monitor when finished running diagnostics on the system
devices.

Note: ADS also recommends performing confirmations
to verify the accuracy of the data and equipment.
Confirmations involve comparing manual flow depth,
temperature, and velocity measurements taken in the field
against monitor readings taken electronically. The
technician or analyst confirms the accuracy of the data
based on the difference between the monitor and field
readings. For more information, refer to the Profile User’s
Guide
(#950015**).

Advertising