ADS Environmental Services FlowShark QR 775002 A4 User Manual

Page 228

Advertising
background image

10-16 ADS FlowShark Manual

Problem

Busy signal occurs when calling the monitor.

Possible

Causes

Someone else may be communicating with monitor.

Monitor may be calling out an alarm.

Telephone cable may be damaged.

Lightning protection module may be damaged.

Telephone cable may have shorted.

Modem in monitor may be damaged.

Telephone service may not be working.

Possible

Solutions

Wait a few minutes, and attempt to communicate with
monitor again.

Connect at the site using the serial cable, and try to
communicate with monitor.

Use voltmeter to check voltage on telephone cable.
Voltage should be approximately 48 Vdc on hook. If it is
not, disconnect phone line at the lightning protection
module and check the voltage at the network interface
box.

Make sure telephone cable is not damaged or severed,
and repair or replace cable if necessary.

Check telephone connector for moisture.

Contact the telephone company to report service is not
working.

Replace the monitor.

Advertising