10-26 ADS FlowShark Manual

Problem

Temperature reading from the pressure depth sensor
shows a value of -273

° C (-459° F).

Possible

Causes

Cable connecting the pressure sensor to the monitor
may be loose or broken.

Pressure sensor may be faulty.

Analog board is not responding or may be faulty.

Possible

Solutions

Secure pressure sensor connection to monitor (if
necessary).

Replace the pressure sensor.

Replace the monitor if defective.

Problem

Pressure depth readings are consistently incorrect.

Possible

Causes

Coefficients may be incorrect.

Pressure dryer tube may not be functioning properly.

Possible

Solutions

Retrieve or enter the correct coefficients, and reactivate
the monitor.

Inspect color of desiccant in dry tube and replace the
dryer tube or desiccant (if necessary).

Problem

Pressure depth readings are consistently off by up to 3
inches, but temperature readings are accurate.

Possible

Cause

Monitor may be configured with an incorrect pressure
sensor offset.

Possible

Solution

Re-measure pressure sensor physical offset and enter
correct value in Profile. Reactivate the monitor.

Enable Auto-Pressure Calibration feature in Profile to
calibrate the pressure sensor data to the ultrasonic data
during non-surcharge conditions.