Maintenance and Troubleshooting 10-21

Problem

Range from ultrasonic depth subsystem is significantly
greater than manually measured range.

Possible

Causes

Temperature or physical offset(s) may be incorrect.

Pipe height may be incorrect.

Sensor may not be level.

Foam or other substance may be absorbing the pulse.

Sensor may be faulty.

Possible

Solutions

Verify the temperature and physical offsets.

Verify the pipe height.

Make sure the sensor is level and in good condition.

Check the hydraulic conditions in the pipe.

Fire the ultrasonic depth sensor at a shorter distance
onto a hard surface to confirm accuracy.

Replace the sensor.

Contact your regional ADS representative.

Problem

Range from the ultrasonic depth subsystem is too short
(but not zero).

Possible

Causes

Temperature or physical offset(s) may be incorrect.

Pipe height may be incorrect.

Ultrasonic depth sensor may be dirty.

Ultrasonic depth sensor may be faulty.

Possible

Solutions

Verify the temperature and physical offsets.

Verify the pipe height.

Clean the ultrasonic depth sensor.

Replace the ultrasonic depth sensor.

Contact your regional ADS representative.