Troubleshooting Guide (continued)

Probable Recommended

Problem

Cause

Solution

2. High background.

a. Exposure time was

i. Decrease

too long.

exposure time.

ii. The use of

Immun-Star
enhancer is
required when
using nitrocellu-
lose membranes.
This will allow
decreased film
exposure times.

b. TBS washes after

i. The washes are

transfer were

critical to reduce

omitted or insufficient.

spotted or
blotchy
background
development.

c. Blocking was

i. Increase the

insufficient.

duration of the
blocking step
and/or the
concentration of
blocker used.

ii. Try a different

blocking reagent,
such as casein,
gelatin, or BSA.