1 preset method programming diagram – Bio-Rad PROTEAN® IEF Cell Accessories User Manual

Page 16

Advertising
background image

4.1 Preset Method Programming Diagram

13

REHYDRATION

>

PRESET METHOD

>

STORED METHOD

>

NEW METHOD

>

SELECT VOLTAGE SLOPE:

RAPID

V

>

LINEAR

V

>

SLOW

V

>

REHYDRATION: yes NO >

GEL LENGTH 07

11 17 >

FOCUS TEMP: 20°

>

NEXT

>

REHYDRATION STEP

active @50V

PASSIVE >

ENTER TEMP: 20 °C >

NEXT

>

ENTER TIME: 12:00

>

INSERT PAUSE AFTER

REHYDRATION? yes NO

>

NEXT

>

METHOD: NAME

LIMIT/ GEL: 50 µA

ENTER # OF GELS:

1 >

PRESS

TO START

REHYDRATION COMPLETE.

PRESS

TO START

STEP #1 OF METHOD.

METHOD IS PAUSED.
EDIT # GELS: >
EDIT METHOD >
PRESS

TO CONTINUE

s 01/04

XXV

XXµA

20/20 °C

XXµA/GEL

VH: 000000

00:00/00:15

LINEAR

V

RUN IN

PROGRESS. (Scrolled
message during run)

REHYDRATION: PASSIVE

(or) AT 50V

TEMP: 20/20 °C

TIME: 00:00/00:00

METHOD WILL START (or
PAUSE) WHEN REHYDRATION
COMPLETE (Scrolled message)

S1

250 V

00:15

S2

HRS:MIN

vhours

>

S2

4000 V

2:00

>

NEXT

>

S3

hrs:min

VHOURS

>

S3

4000 V

20000

>

S4 500 V/HOLD yes NO >

NEXT

>

Continue

To real time edit mode

**CELL COVER OPEN**

POWER SUPPLY PAUSED.

CLOSE COVER

TO PROCEED.

**CELL COVER OPEN**

REHYDRATION PAUSED.

CLOSE COVER

TO PROCEED.

When rehydration is

completed with inserted

pause OR, when pressing STOP

key during rehydration

NO REHYDRATION

selected

REHYDRATION

selected

NEXT if YES

NEXT if NO

Advertising