Fig. 4.24. Positioning glass plates, spacers, and clamps.

4. Grasp the gel sandwich firmly. Guide the left and right clamps onto the sandwich so that

the long and short plates fit the appropriate notches in the clamp (Figure 4.25). Tighten the
screws enough to hold the plates in place.

Fig. 4.25. Attaching the clamps to the glass plate assembly.

5. Place the sandwich assembly in the alignment slot (the slot without cams) of the casting

stand with the short glass plate forward (Figure 4.26). Loosen the sandwich clamps and
insert an alignment card to keep the spacers parallel to the clamps.

Note: Always use the alignment slot and alignment card to set the spacers in place. Failure
to use these can result in gel leakage when casting, as well as buffer leakage during the
run.