Bio-Rad DCode™ Universal Mutation Detection System User Manual

Page 80

Advertising
background image

9.1 Equipment (continued)

Problem

Cause

Solution

Perpendicular gradient (DGGE only)

Glass plate cracked

Excessive force at

Apply only one turn

comb gasket

to thumb gasket screw
after it touches glass

Gel solution leaks during casting

Not sufficient pressure

Make sure pressure

on comb gasket

clamp screws are
turned two turns

Poor contact between

Reassemble glass

comb gasket and

plates as in Section

spacers

4.1. Visually confirm
contact at spacers and
comb gasket

Casting stand gasket

Position gray gasket

positioned incorrectly

so that it covers the entire
bottom section of the
glass plate

Wrong comb gasket

Make sure correct
comb gasket is used

Misaligned comb gasket

Insure that comb gasket
notches are against
spacer notches

Misaligned plates

Pressure clamp may
force plates to shift if
sandwich clamps are not
tight. Insure that
sandwich clamps are
tightened

Misalignment of spacers

Check alignment at

and glass plates

bottom of glass plates,
using alignment slot on
casting stand

Damaged or dirty

Replace spacers or

spacers and/or combs

combs

Different thickness of

Use spacers and comb

spacers and comb

of same thickness

Dirty inlet fitting

Replace fitting

or missing O-ring

Loose stopcock

Tighten stopcock/inlet
port screw connection

No air vent plug

Plug vent after casting gel

Damaged or non-

Replace with Bio-Rad

Bio-Rad glass plates

glass plates only