Bio-Rad DCode™ Universal Mutation Detection System User Manual

Page 85

Advertising
background image

Section 10
Specifications

Construction

Tank, core and clamps

Tank: molded polycarbonate; Core: molded polysulfone;
Clamps: molded glass-filled polycarbonate

Lid

Urethane or polycarbonate

Electrodes

0.010" diameter platinum

Electrical leads

Polyurethane, copper conductor

Casting stand

Able to cast two 16 x 20 cm, two 16 x 16 cm,
two 16 x 10 cm or one 7.5 x 10 cm gels per setup
simultaneously

Heater and control

Temperature control (PID type) ± 0.5°C variation within
gel area, ± 0.5°C actual in the range of 45 to 70°C.Maximum set
temperature 70.5°C

Electrophoresis cooling

Temperature control between 5°C–room temperature with
recommended chillers. Minimum temperature 5°C (SSCP system
only, external chiller required)

DC voltage limit

500 V DC

DC power limit

50 W

Gradient former

Cast acrylic and acetal (DGGE system only)

Glass plates

20 x 22 cm (20 cm format), 16 x 20 cm (16 cm format),
10.2 x 20 cm (10 cm format)

Gel sizes

16 x 20 cm (max. two per run), 16 x 16 cm (max. two per run),
16 x 10 cm (max. two per run), 7.5 x 10 cm (max. four per run)

Spacers available

0.75, 1.0 and 1.5 mm

Combs

16 well comb (compatible with 8 well multi-channel
pipettor), 20 well comb, 25 well comb and 1 well comb
(Prep comb for perpendicular gradient gels). Optional can
use combs from PROTEAN

®

II xi system

AC Power Requirements

170-9080/9083/9086/9089

AC power input: 120 VAC 47–63 Hz, 5 A slow blow fuse

170-9092/9095/9098/9102

170-9082/9085/9088/9091

AC power input: 100 VAC 47–63 Hz, 5 A slow blow fuse

170-9094/9097/9100/9104

170-9081/9084/9087/9090

AC power input: 220–240 VAC 47–63 Hz, 2.5 A slow blow fuse

170-9093/9096/9099/9103

DC Power Requirements

External DC voltage power supply. This power supply must be ground isolated in such a way that the
DC voltage output floats with respect to ground.

Maximum voltage limit

500 VDC

Maximum power limit

50 W

Size and Weight

Overall size

Lid and tank assembly: 39 cm (L) x 20 cm (W) x 42 cm (H)

Shipping weight

16 Kg

Environmental Requirements

Storage environment

0–70°C, humidity 0–95% (non-condensing)

Operating environment

0–35°C, humidity 0–95%

Regulatory

Meets requirements of EN61010-1.