Analysis tool box tools – Bio-Rad Image Lab™ Software User Manual
Page 129

Analysis Tool Box Tools
User Guide | 129
The band detection sensitivity options are as follows:
Low Band Detection Sensitivity — sets detection at a low level for images with
prominent bands. Faint bands are not detected with this setting.
High Band Detection Sensitivity — sets detection at a higher level for images that
are faint. Extraneous bands can be removed using the Band Tools in the Analysis
Lane and Bands Tools on page 133
Custom — enables you to set a value between 1 and 100 to select the best
detection sensitivity for your sample. You can also drag the sliding bar left or right to
set the value.
When Low Band Detection Sensitivity is used, the numerical value is set at 25; when
High Band Detection Sensitivity is used, the value is set at 75.
Molecular Weight Analysis Settings
Molecular Weight Standard — choose any of the many Bio-Rad standards or
another standard you added to your standards list.
Standard Lanes — choose or change the lanes in which the standards are placed.
Regression Method — four regression methods are available. For more
Regression Calculation Methods on page 253
Analysis Tool Box Tools
All Analysis Tool Box tools customize the analyzed data in image files. These tools
are available only when an image file is open. Click a specific image to select among
the many windows that might be open in your workspace.
Note: Some tools delete the existing analysis.
To access a tool
Click any of the toolbox buttons.