Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 15

Advertising
background image

2

3

Bio

8.

Enter user name and control number. Select Optics. Select OK. The
following dialog appears: (Figure 4).

Fig. 4.

9.

Select the Eject button in the dialog box to eject the plate holder.

10. Place the MCV plate II in the microplate platform.

11. Select OK to begin the optics validation procedure.

12. When the procedure has been completed, the results will be displayed in a

dialog box as shown in Figure 5.

13. The results will also be logged into a validation log in Bio-Plex Manager. To

access this log, select View from the main menu then Validation Log. Each
type of validation is stored in a separate log for the purpose of tracking data
over time. See Bio-Plex Manager 3.0 user manual for more information on
using the validation log.

Bio-Plex Manager 3.0 and MCV Plate II Instructions

12

4110007c.qxd 02/25/2003 10:12 AM Page 12